Development and validation of a liquid chromatography-isotope dilution tandem mass spectrometry for determination of olanzapine in human plasma and its application to bioavailability study / 药学学报
Yao Xue Xue Bao ; (12): 767-71, 2010.
Article in En | WPRIM | ID: wpr-382484
Responsible library: WPRO
ABSTRACT
A simple, reliable and sensitive liquid chromatography-isotope dilution mass spectrometry (LC-ID/MS) was developed and validated for quantification of olanzapine in human plasma. Plasma samples (50 microL) were extracted with tert-butyl methyl ether and isotope-labeled internal standard (olanzapine-D3) was used. The chromatographic separation was performed on XBridge Shield RP 18 (100 mm x 2.1 mm, 3.5 microm, Waters). An isocratic program was used at a flow rate of 0.4 m x min(-1) with mobile phase consisting of acetonitrile and ammonium buffer (pH 8). The protonated ions of analytes were detected in positive ionization by multiple reactions monitoring (MRM) mode. The plasma method, with a lower limit of quantification (LLOQ) of 0.1 ng x mL(-1), demonstrated good linearity over a range of 0.1 - 30 ng x mL(-1) of olanzapine. Specificity, linearity, accuracy, precision, recovery, matrix effect and stability were evaluated during method validation. The validated method was successfully applied to analyzing human plasma samples in bioavailability study.
Full text: 1 Index: WPRIM Language: En Journal: Yao Xue Xue Bao Year: 2010 Type: Article