Your browser doesn't support javascript.
loading
The effect of X-ray radiography exposure intensity on image quality of the monitor / 中华放射学杂志
Chinese Journal of Radiology ; (12): 1056-1058, 2008.
Article in Chinese | WPRIM | ID: wpr-398345
ABSTRACT
Objective To explore the relationship between image quality and different exposure intensity of digital radiography(DR).Methods Using a contrast-detail phantom Cdrad 2.0,observers evaluated images on the monitots when exposure parameters were 90 kV and mAs ranged from 0.5 to 16.0 mAs.Then calculate image quality figure(IQF),and ANOVA was used to analyze the difference of IQF under different radiation dose in 8 subjects.Results (1)When exposure parameters were 90 kV and mAs altered from 0.5 to 8.0 mAs.the IQF observed on the monitor was reduced,and the minimal value was 38.54±2.82.When mAs changing from 8.0 to 16.0 mAs.IQF observed on the monitor was increased,and the maximum was 44.33±2.35.(2)Under the different exposure condition,the difference of IQF values were statistically significant(F=60.35,P<0.01).Conclusions With the increased exposure intensity,the image quality on monitor increased.When mAs was 8.0 mAs,image quality was the optimum When mAs was more than 8.0 mAs,the image quality on monitor reduced.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiology Year: 2008 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiology Year: 2008 Type: Article