Your browser doesn't support javascript.
loading
Dosimetric evaluation of carbon fiber tabletop on absorbed doses / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 223-225, 2008.
Article in Chinese | WPRIM | ID: wpr-401477
ABSTRACT
Objective To evaluate the dosimetric effect of carbon fiber tabletop on the patient doses in radiotherapy. Methods The transmission factors of couch and inserts were measured in air and solid water phantom using 0.6 cm3 ion chamber and PTW 2D ion chamber array for 6 MV ,10 MV and 18 MV X-ray,respectively. Absorbed doses at depth of maximum dose ,5 cm and 10 cm in solid water were measured with the 2D ion chamber array. Absorbed doses fluctuations with different gantry angles and air gaps between phantom and couch were also measured. Results The posterior field measurement showed that the reductions of absorbed doses at the depth maximum dose,5 cm and 10 cm were within 5%. The ratios of the absorbed doses with to without couch increased with the oblique incident angles and varied slightly with the air gap at depth of 5 cm. The transmission factors of inserts were less than those of couch owing to its thinner thickness. Conclusion The carbon-fiber tabletop affects the absorbed doses and dose distributions of the target, and this effect changes with the gantry angle and air gap. Special considerations should be taken during treatment planning.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2008 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2008 Type: Article