Patient Setup Aid with Wireless CCTV System in Radiation Therapy / 대한방사선종양학회지
The Journal of the Korean Society for Therapeutic Radiology and Oncology
;
: 300-308, 2006.
Article
in Korean
| WPRIM
| ID: wpr-40222
ABSTRACT
PURPOSE:
To develop a wireless CCTV system in semi-beam's eye view (BEV) to monitor daily patient setup in radiation therapy. MATERIALS ANDMETHODS:
In order to get patient images in semi-BEV, CCTV cameras are installed in a custom-made acrylic applicator below the treatment head of a linear accelerator. The images from the cameras are transmitted via radio frequency signal (~2.4 GHz and 10 mW RF output). An expected problem with this system is radio frequency interference, which is solved utilizing RF shielding with Cu foils and median filtering software. The images are analyzed by our custom-made software. In the software, three anatomical landmarks in the patient surface are indicated by a user, then automatically the 3 dimensional structures are obtained and registered by utilizing a localization procedure consisting mainly of stereo matching algorithm and Gauss-Newton optimization. This algorithm is applied to phantom images to investigate the setup accuracy. Respiratory gating system is also researched with real-time image processing. A line-laser marker projected on a patient's surface is extracted by binary image processing and the breath pattern is calculated and displayed in real-time.RESULTS:
More than 80% of the camera noises from the linear accelerator are eliminated by wrapping the camera with copper foils. The accuracy of the localization procedure is found to be on the order of 1.5+/-0.7 mm with a point phantom and sub-millimeters and degrees with a custom-made head/neck phantom. With line-laser marker, real-time respiratory monitoring is possible in the delay time of ~0.17 sec.CONCLUSION:
The wireless CCTV camera system is the novel tool which can monitor daily patient setups. The feasibility of respiratory gating system with the wireless CCTV is hopeful.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Particle Accelerators
/
Copper
/
Hope
/
Head
/
Noise
Limits:
Humans
Language:
Korean
Journal:
The Journal of the Korean Society for Therapeutic Radiology and Oncology
Year:
2006
Type:
Article
Similar
MEDLINE
...
LILACS
LIS