AIM:
To establish HPLC method for the determination of amodiaquine in human plasma.METHODS:
Amodiaquine and internal standard(hydroxychloroquine) were analyzed on C18 column(150 mm ×4.6 mm, 5 μm) with methanol watertriethylamine orthophosphoric acid (2177.510.5 )as mobile phase at the flow rate of 1.0 mL · min-1The UV detector was set at 294 nm.RESULTS:
The retention times of amodiaquine and internal standard were 5.82, 8.56 min, respectively. The calibration curve was linear in the range from 10 to 1 000 μg ·L- 1 ( r = 0.999 8, n = 9 ). The limit of quantitation was5 μg· L-1. The extraction recovery was between 75.5 % and 82.7 %, and the methodological recovery was between 97.0 % and 104.8 %. The intra-day and inter-day RSD were less than 6.0 % and 7.5 %, respectively.CONCLUSION:
This HPLC method is simple, sensitive and suitable for pharmacokinetic study of amodiaquine.