Your browser doesn't support javascript.
loading
Determination of Cyanogen Chloride in the Air Pollution by Spectrophotometry / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 565-568, 2001.
Article in Chinese | WPRIM | ID: wpr-410356
ABSTRACT
We report a spectrophotometic method for the determination of cyanogen chloride (CNCl) in air pollution.Experimental parameters for the stability and absorbility of CNCl gas in the different media were studied.The experimental conditions were optimized for CNCI as followsabsorbent1%isonicotinic acid-barbituric acid;reaction aciditypH 5.8 in phosphate buffer at room temperature;masking agent0.01mol/L EDTA;λmax598nm. The apparent molar absorptivity was found to be 1.17×105L·mol-1·cm-1,the linear range was within 0~5μg of CNCl in 25 mL solution; the regression equation of the curve has given by A=-0.0040+1.935C(r=0.9999).The method has been applied to the determination of CNCl in chimney gas,workshop air and production area air of the cyanuric trichloride plant,chemical works and pharmaceutical factory air.The relative standard deviation was below 2.7%(n=4)for the cyanogen chloride amount within 13~64mg/m3,and when CNCl amounts were within 0.1~1.0mg/m3,RSD was below 6.9%(n=6).The recovery was 98.7%.
Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Analytical Chemistry Year: 2001 Type: Article