Experiment of Detecting Total Dissolved Solids in Drinking Water Using Conical Flask Method / 环境与健康杂志
Journal of Environment and Health ; (12): 366-367, 2000.
Article in Chinese | WPRIM | ID: wpr-412242
ABSTRACT
[Objective] To explor the possibility of application of conical flask as an evaperation vessel in detecting the content of total dissolved solids in drinking water. [Methods]The conical flask was placed on a electric stoveto dry, and in electric oven to bake until the weight of conical flask was constant, at temperature of 180±3℃, andthen was used to detect the total dissolved solids in drinking water. The comparison between conical flask methodand National Standard Method was made. [Results] The levels of total dissolved solids obtained according to conical-flask method and National Standard Method revealed no significaut differences (P>0.5). [Conclusion] The detection of total dissolved solids in drinking water by conical flask method was rapid, accurate and practicabel.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Environment and Health Year: 2000 Type: Article