Evaluation and test of 192Ir air kerma strength for afterloading systems / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 76-78, 2011.
Article in Chinese | WPRIM | ID: wpr-414046
ABSTRACT
Objective To study the method of measuring air kerma strength of afterloading units with 192Ir source by using well type ionization chamber.MethodsThe air kerma strength of 30 afterloading units with 192Ir source was measured using 2000A electrometer and 1000 plus well type ionization chamber,and apparent activity of the source was calculated with the air kerma strength and apparent activity conversion factor.The measured activity of the source was compared with the original value of the source provided by the manufacturer,and the relevant deviation should be within ± 5%.Results Theair kerma strength of afterloding units with 192Ir sources was tested.The relevant deviation of the measured activity and the original value was within -0.1%-4.4%.Conclusions The measurement method with a well type ionization chamber is convenient and highly accurate which can be used for the test of quality control in hospitals.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2011 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2011 Type: Article