Your browser doesn't support javascript.
loading
Human reliability in high dose rate afterloading radiotherapy based on FMECA / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 314-317, 2012.
Article in Chinese | WPRIM | ID: wpr-427084
ABSTRACT
Objective To put forward reasonable and feasible recommendations against the procedure with relative high risk during the high dose rate (HDR) afterloading radiotherapy,so as to enhance its clinical application safety,through studying the human reliability in the process of carrying out the HDR afterloading radiotherapy.Methods Basic data were collected by on-site investigation and process analysis as well as expert evaluation.Failure mode,effect and criticality analysis (FMECA) employed to study the human reliability in the execution of HDR afierloading radiotherapy.Results The FMECA model of human reliability for HDR afterloading radiotherapy was established,through which 25 procedures with relative high risk index were found,accounting for 14.1% of total 177 procedures.Conclusions FMECA method in human reliability study for HDR afterloading radiotherapy is feasible.The countermeasures are put forward to reduce the human error,so as to provide important basis for enhancing clinical application safety of HDR afterloading radiotherapy.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2012 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2012 Type: Article