Establishment of daily quality assurance program and analysis of dosimetric characteristics’ long- term stability for mobile intraoperative radiation therapy accelerator / 中国癌症杂志
China Oncology ; (12): 52-56, 2014.
Article in Chinese | WPRIM | ID: wpr-439515
ABSTRACT
Background and

purpose:

Currently, intraoperative radiation therapy (IORT) has become the adjuvant therapy of cancer. The study was to establish the daily quality assurance (QA) program and analyze dosimetric characteristics’ long-term stability for mobile IORT accelerator.

Methods:

The QA program of this study included two partssafety and functionality and energy and output. The two years’ QA datasets were acquired and analyzed to investigate the stability of energy and output.

Results:

All safety and functionality tests passed on a daily basis. The energy index was (0.666±0.015)mm, (0.839±0.009)mm, (0.781±0.010)mm, (0.724±0.009)mm and the output dose error was (0.511 ± 0.671)%, (0.278 ± 0.516)%, (0.368 ± 0.532)%, (0.382 ± 0.912)%for all energy, respectively. There was no signiifcant time trend in the dosimetric characteristics.

Conclusion:

The daily QA program is suitable for mobile IORT accelerator. The long-term stability is acceptable for IORT in clinical use.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Oncology Year: 2014 Type: Article