Your browser doesn't support javascript.
loading
Analysis of set-up accuracy of two body membrane fixation methods in radiotherapy for thoracic neoplasms / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 397-399, 2013.
Article in Chinese | WPRIM | ID: wpr-440484
ABSTRACT
Objective To study the set-up accuracy in radiotherapy of thoracic neoplasms by improving the body immobilization method.Methods Fifty patients with thoracic neoplasms were randomly divided into conventional group (without electrode paste) and improved group (with electrode paste).Using simulator for position calibration and center field digital image reconstruction from treatment planning system.Then compare the set-up accuracy of two groups with different body methods by grouped t-test.Results Set-up error in the left-right,superior-inferior,anterior-posterior direction were 2.5 ± 1.5 and 2.4 ± 1.4(P =0.010),4.4 ± 2.0 and 2.2 ± 1.2 (P =0.000),2.2 ± 1.3 and 2.1 ± 1.0 (P =0.100) in conventional group and improved group,respectively.Conclusions The improved body immobilization method improves setup accuracy in radiotherapy for thoracic neoplasms which also will be effective for clinical treatment.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2013 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2013 Type: Article