ABSTRACT
Graphene/Room temperature ionic liquids ( GN/IL ) nanocomposite was prepared by grinding graphene and ionic liquid 1-butyl-3-methylimidazolium bromide ( BMIMPF6 ) which mixed together in appropriate proportion. Atomic force microscopy ( AFM ) was utilized to characterize the formation of the GN-ILs. Due to the synergistic effect between ionic liquids and graphene, the nanocomposite exhibited excellent performance toward H2 O2 reduction. A novel uric acid ( UA) electrochemical sensor was fabricated based on uricase-GN/IL modified glassy carbon electrode. The experimental results showed that the response displayed a good linear response toward UA in the concentration range from 0. 002-4. 5 mmol/L. The corre1ation coefficient was 0. 995 and the detection limit was 0. 85 μmol/L. The easily prepared electrochemical sensor had favorable stability and selectivity and could be applied to the quick determination of UA in human serum, thus providing a new UA detection method for clinical trial.