Clinical application of IMSure QA software / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection
;
(12): 292-294, 2014.
Article
in Chinese
| WPRIM
| ID: wpr-446659
ABSTRACT
Objective To compare IMSure QA and MatriXX for intensity modulated radiotherapy plans,and to investigate the credibility of IMSure QA software.Methods Ten patients receiving intensity modulated radiotherapy were randomly chosen for the quality assurance plans with IMSure QA and MatriXX.Results The overall average of percentage pass points in 3% and 3 mm were (98.1 ± 0.8) % with IMSure QA,and (97.9 ±0.6)% with MatriXX(t =0.86,P >0.05).Conclusion IMSure QA can be a reliable verification tool for IMRT QA.