Peak skin dose measurements in two cardiac interventional procedures using radiochromic film / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 302-305, 2014.
Article in Chinese | WPRIM | ID: wpr-446660
ABSTRACT
Objective To measure the peak skin dose (PSD) in two cardiovascular interventional procedures,including coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) using radiochromic film.Methods Gafchromic XR-RV3 film was selected to measure PSD in two hospitals.The films were placed on the table underneath the patient during interventional surgery.The kV,mA,fluoroscopy time,dose-area product (DAP),and cumulative dose at reference point and other relevant information were recorded for all cases.Using the Epson V750 flatbed scanner for scanning and analyzing film,FilmQA software was chosen to analyze the pixel value of red,green and blue color channels.The PSD was determined using red channel data.The correlation and linear regression analysis between PSD and device-displayed parameters was carried out.Results PSD were measured using XR-RV3 film for 26 CA and 19 CA + PTCA procedures.For CA procedures,maximum fluoroscopy time,cumulative dose and DAP were 17.62 min,1 498.50 mGy and 109.68 Gy · cm2,respectively.The maximum PSD was 361.20 mGy.However,for CA + PTCA procedures,maximum fluoroscopy time,cumulative dose and DAP were 64.48 min,6 976.20 mGy and 5 336.00 Gy· cm2,respectively.One patient with CA + PTCA procedures was found to have received the PSD value more than 2 Gy,up to 2 195.70 mGy.DAP was found to be a good indicator (R2 =0.815,P <0.05) of PSD for CA procedure,and correlated with cumulative dose (R2 =0.916,P < 0.05) for CA + PTCA procedures.Conclusions The PSD value of some patients in cardiac interventional procedures would exceed 2 Gy,the threshold of deterministic effects recommended by ICRP.The dose-related parameters value showed on DSA device can only used to estimate PSD roughly.Using XR-RV3 film accurate measurement of the PSD in interventional projects is a very fast and effective method.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2014 Type: Article