Your browser doesn't support javascript.
loading
Comparative analysis on blood glucose results detected by portable glucose meter and Olympus AU2700 automatic biochemical analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 1768-1769, 2014.
Article in Chinese | WPRIM | ID: wpr-451920
ABSTRACT
Objective To perform the comparative analysis on the blood glucose results detected by the portable glucose meter and the Olympus AU2700 biochemical analyzer.Methods 20 fresh whole blood samples with sodium fluoride and potassium oxalate anticoagulation were the Olympus AU2700 biochemical analyzer biochemical analyzer.Results In these 27 portable glucose meters, the comparison results in only 1 portable glucose meter was unqualified,which in the other 26 portable glucose meters were passed, the passing rate was 96.3%,which in 2 glucose meters were clinically acceptable and which in the other 24 glucose meters were completely acceptable and the clinical completely acceptable rate was 88.9%.Conclusion There are some biases of blood glucose results detected by the portable glucose meter and the fully automatic biochemical analyzer,but the overall correlation is well.The comparison and calibration of the blood glucose meters should be performed at the regular intervals for improving the consistency and accuracy of blood glucose detection results.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2014 Type: Article