Study on inter-accreditation of electrolytes detection results in grade 3A hospitals of Zhuhai city / 国际检验医学杂志
International Journal of Laboratory Medicine
;
(12): 2073-2074, 2014.
Article
in Chinese
| WPRIM
| ID: wpr-455162
ABSTRACT
Objective To provide the scientific basis for realizing the inter-accreditation of laboratory electrolyte detection results by comparing the electrolyte detection results in 5 grade 3A hospitals of Zhuhai city .Methods Each 10 serum samples with low , middle and high concentrations of electrolyte were collected for simultaneously detecting the electrolyte kalium (K) ,natrium (Na) and chlorinum (Cl) .The detection results were performed the statistical analysis and comparison .The mutual bias within 1/2 of al-lowable error of CLIA′88 indicated that the detection results were mutually accredited ,if the mutual bias exceeding 1/2 of allowable error ,the detection results could not be mutually accredited .Results The difference of electrolyte detection results in 5 hospitals accorded with the stipulation requirement of CLIA′88 .Conclusion The electrolyte detection results of 5 hospitals could be mutually accredited .
Full text:
Available
Index:
WPRIM (Western Pacific)
Type of study:
Diagnostic study
Language:
Chinese
Journal:
International Journal of Laboratory Medicine
Year:
2014
Type:
Article
Similar
MEDLINE
...
LILACS
LIS