Determination of Residual Solvent DMF in Imiquimod by HS-GC / 中国药师
China Pharmacist ; (12): 1586-1588, 2014.
Article in Chinese | WPRIM | ID: wpr-456288
ABSTRACT

Objective:

To establish a new method for determining residual organic solvent DMF in imiquimod.

Methods:

The samples were injected into an Agilent HP-PLOT/Q capillary column(30 m × 0. 530 mm,40. 0 μm) by a headspace sampler and ana-lyzed with an FID detector, the carrier gas was nitrogen, the inlet temperature was 250℃, and the detector temperature was 270℃. The column temperature was programmed raised.

Results:

The resolution among the peaks of DMF and the other residual solvents could meet the requirements. There was a good linearity within the experimental concentration range. The average recovery was 94. 6%(RSD=4. 0%, n=9). The limit of quantification and the limit of detection was 4. 809μg·ml-1 and 0. 963μg·ml-1, respectively.

Conclusion:

The method is convenient, accurate and sensitive, which can be used in the determination of residual solvent DMF in imi-quimod.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Pharmacist Year: 2014 Type: Article