Your browser doesn't support javascript.
loading
The comparative analysis of automatic blood analyzer Sysmex XT-2000i and Mindray BC-5800 / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 2372-2373, 2014.
Article in Chinese | WPRIM | ID: wpr-456509
ABSTRACT
Objective To compare the precision and correlation of test parameters betwee the two types of automatic blood ana-lyzer,Sysmex XT-2000i and Mindray BC-5800.Methods We selected 60 cases of EDTA-K2 anticoagulant whole blood samples, correlation analysis was carried out on the two instrument parameters,and high,medium and low value samples were chosen for 10 times′precision analysis in the two instrument s respectively.Results The variation coefficient(CV)values of blood count parame-ters of Sysmex XT-2000i were 0.02%-5.70%,while the CV values of white blood cell classification parameters were 0.06%-1. 66%.The CV values of blood count parameters of Mindray BC-5800 were 0.02%-5.3%,white the CV values of white blood cell classification parameters were 0.09%-1.17%.The determination coefficient(r2 )of each parameter of the two instruments were 0. 185-0.995.Conclusion the correlation between the two instruments is good,and the two instruments′precisions were consistent with the design range.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2014 Type: Article