Your browser doesn't support javascript.
loading
The study of error in the head and neck phantom using the X-ray volumetric image system of Elekta Synergy accelerator with CBCT / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 433-436, 2014.
Article in Chinese | WPRIM | ID: wpr-457020
ABSTRACT
Objective To investigate the accuracy and reliability of the image automatic matching using X-ray volumetric image (ⅩⅥ) system with the accelerator Elekta Synergy and to compare the matches with different methods and area.Methods A simulation plan using the head and neck phantom was designed and sent to ⅩⅥ system for CBCT scanning.During the scanning,the couch for certain distance was moved and the data of ⅩⅥ image automatic Matching in different matching methods and area was collected.Results The result of ⅩⅥ image automatic matching is consistent and correct to direction and distance which the couch had been moved (x(0.11 ± 0.41) mm,y(-0.04± 2.6) mm,z(0.28 ± 0.74 mm)).There are not significant differences between the results that matching in different matching methods and different matching area.Conclusions The automatic image Matching of ⅩⅥ is accuracy and high reliability in recognition of offset error.But there are some significant differences on the automatic image Matching in different matching methods and different matching area.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 2014 Type: Article