Development of a QuEChERs Method for Determination of 20 Perfluorinated Compounds in Animal Liver by HPLC_MS/MS / 分析化学
Chinese Journal of Analytical Chemistry
;
(12): 40-48, 2015.
Article
in Chinese
| WPRIM
| ID: wpr-457746
ABSTRACT
A method for simultaneous determination of 20 Perfluorinated alkyl substances ( PFAS) in animal liver using QuEChERs and HPLC_MS/MS technique was developed. The samples were extracted with 0. 1%hydrochloric acetonitrile and cleaned up with C18 , N_Propylethylendiamine ( PSA ) and graphitized carbon blacks ( GCB ) . The analytes were separated by a reversed phase C18 column and gradiently eluted with a mixed solution of 5 mmol/L ammonium acetate methanol and 5 mmol/L ammonium acetate. The samples were quantified using isotope internal standard and external standard with the matrix matched standard calibration curve method. Good linearity was obtained for all the 20 PFAS at the concentration of 0. 1-10 μg/L with the linear correlation coefficients more than 0. 9995. The limits of detection (LOD) and the limits of quantification ( LOQ) for PFAS were 0. 05-0. 2 μg/kg and 0. 4-0. 5 μg/kg, respectively. The recoveries at three different concentration levels ( 0 . 5 , 2 and 5 μg/kg ) were in the range of 70 . 3% -108 . 1%. The repeatability expressed as relative standard deviations (RSD) was ranged from 2. 1% to 11. 9% (n=6).
Full text:
Available
Index:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Chinese Journal of Analytical Chemistry
Year:
2015
Type:
Article
Similar
MEDLINE
...
LILACS
LIS