Objective To compare the performance of high‐throughput ELISA and ECLIA in the determination of carcinoembry‐onic antigen (CEA) .Methods The CEA concentration of serum samples were respectively determinate by high‐throughput ELISA and ECLIA ,and the results were compared .Results Two kinds of detection methods could both accurately reflect the concentration of serum CEA .There was no significant difference between the results of two methods (P> 0 .05) .Within the linear range ,the CEA result of high‐throughput was correlate closely with that of ECLIA (r=0 .922 8 ,P<0 .01) .Conclusion High‐throughput ELISA can accurately detect the serum CEA concentration .