Your browser doesn't support javascript.
loading
Study on results comparability of detecting tumor markers by two kinds of chemiluminescence system / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 3251-3252, 2014.
Article in Chinese | WPRIM | ID: wpr-458490
ABSTRACT
Objective To study the results comparability of two kinds of different chemiluminescence systems(Antu chemilumi-nescence analyzer ADC CLIA 400 and Siemens centaur xp)in the detection of tumor markers.Methods The tumor markers were detected in 50 specimens by using the Antu chemiluminescence analyzer ADC CLIA 400 and the Siemens centaur xp chemilumines-cence analyzer respectively.Then the results were analyzed and compared.Results The quantitative analysis results on the serum specimens showed that there was no significant difference in the detection results between the two kinds of detection system(P >0.05).The correlation coefficients of 5 test items between the two kinds of detection system were more than 0.836,suggesting a good correlation between the two kinds of method.The accuracy and repeatability of the Siemens centaur xp chemiluminescence ana-lyzer were higher than those of the Antu chemiluminescence analyzer ADC CLIA 400 in detecting the tumor markers.Conclusion The Siemens centaur xp chemiluminescence analyzer is better than the Antu chemiluminescence analyzer ADC CLIA 400 in detec-ting the tumor immune markers.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2014 Type: Article