Performance verification for Olympus Au5400 in automatic biochemistry analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 498-500, 2015.
Article in Chinese | WPRIM | ID: wpr-462127
ABSTRACT
Objective To verify and confirm the optical performance of the Olympus AU5400 Automatic Analyzer is satisfied for the professional requirements.Methods According to the Medical Standard of the People's Republic of ChinaAutomatic chemistry analyzer(YY/T 0654-2008),the Analyzer Modular Ⅰand Ⅱ optical performance were measured to verify the stray light,the ab-sorption accuracy,linear range,stability and repeatability.Results The stray light absorption of tow modular was 4.789 4-5.158 9;The absorption accuracy of potassium dichromate standard solution (1-1)was (-0.002 1--0.024 7),and solution (1-2)was (-0.008 6--0.067 7);The D-value of stability that the highest absorption of standard solution (1-1)minus the lowest was 0.000 5-0.001 5;The CV% of repeatability for standard solution (1-2)was 0.02%-0.04%;Therof absorption linear was 0.999 6-0.999 9. Conclusion The optical performance of Olympus AU5400 Analyzer conforms to the standard of YY/T 0654-2008.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2015 Type: Article