Application of moving average method in the quality control of coagulation function measurement / 国际检验医学杂志
International Journal of Laboratory Medicine
;
(12): 1400-1401, 2015.
Article
in Chinese
| WPRIM
| ID: wpr-465016
ABSTRACT
Objective To discuss the feasibility of moving average method (X‐B method) in the quality control of coagulation function measurement .Methods Fluctuation average value tendency of prothrombin time (PT ) ,activated partial thromboplastin time (APTT) ,thrombin time (TT) and fibrinogen (FIB) of clinical specimen ,detected by using Stago Compact analyzes ,were ana‐lyzed ,and compared cryochemic plasma method .Results Average value of PT ,APTT ,TT and FIB changed in small ranges .Tend‐ency of quality control chart of X‐B method and cryochemic plasma method were basically consistent .Conclusion X‐B method ,u‐sing plasma of patients ,could be used for the quality control of PT ,APTT ,TT and FIB examination and for the monitoring of the stability of analyzers .