Low-dose radiation in the hip digital tomosynthesis / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection
;
(12): 153-156, 2015.
Article
in Chinese
| WPRIM
| ID: wpr-466252
ABSTRACT
Objective To explore the optimal dose ratio set in the hip digital tomosynthesis (DTS),and to reduce patient's examination dose as low as possible.Methods Ninety patients who received hip DTS examination were randomly divided into 3 groups,with the dose ratio as 6,7 and 8,respectively.One-way ANOVA test was used to compare.the image quality and radiation dose among 3 groups.Results The values of entrance surface dose (ESD),dose-area product (DAP) and image quality score of 3 groups were [(3.76 ± 1.89) mGy,(18.41 ± 11.71) dGy·cm2,3.03 ± 0.24],[(5.24±2.76)mGy,(26.99±13.34)dGy·cm2),3.60±0.11],and [(6.39±1.75)mGy,(36.96± 22.49) dGy· cm2,3.64 ± 0.09],respectively.The difference among three groups was statistically significant (F =10.94,9.45,139.26,P < 0.05).The S-N-K test showed that both ESD values and DAP values in 6,7,and 8 time-dose ratio group was from small to large.The image quality score of 6 time-dose ratio group was lower than that of the other groups with no significant difference.Conclusions Dose ratio is one of the important parameters to result in the radiation dose of DTS.Dose ration 7 is optimal in hip DTS,which could match the image quality and radiation dose best and finally protect patients from unnecessary radiation damage.
Full text:
Available
Index:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Chinese Journal of Radiological Medicine and Protection
Year:
2015
Type:
Article
Similar
MEDLINE
...
LILACS
LIS