Your browser doesn't support javascript.
loading
Design of a New Applicator for High-Dose Rate Vaginal Brachytherapy / 영남의대학술지
Yeungnam University Journal of Medicine ; : 123-128, 2000.
Article in Korean | WPRIM | ID: wpr-46738
ABSTRACT

PURPOSE:

This study was aimed to develop a new vaginal applicator(Shin's Applicator) for 2-channel high-dose rate vaginal brachytherapy to evaluate uniformity of surface dose, and to present 3-dimensional dose distribution of the applicator.

METHODS:

Shin's Applicator was inexpensively constructed using human soft tissue equivalent acrylic bar. We evaluated dose uniformity along the applicator surface using film densitometer and performed vaginal intracavitary brachytherapy after insertion of the applicator using HDR brachytherapy planning software and brachytherapy unit(Ralstron-20B).

RESULTS:

Shin's Applicator allows improved dose distribution than the existing 1-channel cylinder and achieves diminished urinary bladder and rectal dose by 20%.

CONCLUSIONS:

From the above results, it can be concluded that Shin's Applicator may be an improved form of a vaginal applicator. Furthermore, it can be suggested that this applicator has an advantage, for it prevents vaginal stenosis after radiation therapy and can be used as a disposable vaginal dilator. Further follow up examination with radiological study may be helpful to evaluate the therapeutic efficacy of this applicator.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Urinary Bladder / Brachytherapy / Constriction, Pathologic Limits: Humans Language: Korean Journal: Yeungnam University Journal of Medicine Year: 2000 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Urinary Bladder / Brachytherapy / Constriction, Pathologic Limits: Humans Language: Korean Journal: Yeungnam University Journal of Medicine Year: 2000 Type: Article