Comparison of periphery capillary whole blood glucose using four kinds of blood glucose meters and venous plasma glucose using laboratory auto-analyzer / 中国基层医药
Chinese Journal of Primary Medicine and Pharmacy
;
(12): 582-585, 2015.
Article
in Chinese
| WPRIM
| ID: wpr-470421
ABSTRACT
Objective To evaluate the precision of the ACCU-CHEK Performa glucose monitor,the Optium Xceed glucose monitor,the Breeze glucose monitor and the Contour glucose monitor.Methods A total of 102 patients (diabetic patients or un-diabetic patients) who visited our endocrinology outpatient department were randomly selected.The venous blood samples were collected using the laboratory auto-analyzer,and the synchronous finger tip capillary blood were collected using the ACCU-CHEK Performa glucose monitor and the Optium Xceed glucose monitor,or using the Breeze glucose monitor and the Contour glucose monitor.Results The correlation coefficients between VPG by the laboratory auto-analyzer and CBG by the four kinds of glucose monitors was good,and R values were 0.990,0.985,0.963,0.952,and there was no significant difference between the CBG of Breeze glucose monitor and the same VPG(P > 0.05).When the blood glucose concentration was higher than 4.2mmol/L,and compared with the VPG detected by synchronous measurement laboratory,95% of the CBG test results detected by four kinds of blood glucose meters were in the range of ± 20%.Conclusion The four kinds of glucose monitors can provide high accurate.