Your browser doesn't support javascript.
loading
Reproducibility of normal hepatic proton MRS at 3.0T / 中国医学影像技术
Article in Zh | WPRIM | ID: wpr-472945
Responsible library: WPRO
ABSTRACT
Objective To explore the reproducibility of normal hepatic MRS at 3.0T. Methods The hepatic proton MRS was performed with GE Signa Excite HD 3.0T system and eight-channel torso phased-array coils using PRESS sequence. Thirty-one healthy individuals were enrolled in this study. Liver spectra were collected with TR of 1500 ms, TE of 30 ms, ROI of 2 cm×2 cm×2 cm, NSA of 64 times. The outcomes were statistically analyzed with Wilcoxon signed ranks test and Spearman correlation test.Results There was no significant difference of signal to noise ratio (Z=-0.535,P=0.593), baseline stability (Z=-0.333, P=0.739), linewidth of automatic shimming (Z=-0.305, P=0.761), water suppression (Z=-1.232, P=0.218), height of lipid peak (Z=-0.558,P=0.557), area under the lipid peak (Z=-1.195,P=0.232), height of water peak (Z=-0.647,P=0.518) and area under the warter peak (Z=-0.118, P=0.906) between first examination and second examination. Correlation coefficient of the former and the later measurements of lipid area and water area were 0.784 (P<0.001) and 0.799 (P<0.001), respectively.Conclusion The reproducibility is good for in vivo liver proton MRS at 3.0T.
Key words
Full text: 1 Index: WPRIM Language: Zh Journal: Chinese Journal of Medical Imaging Technology Year: 2009 Type: Article
Full text: 1 Index: WPRIM Language: Zh Journal: Chinese Journal of Medical Imaging Technology Year: 2009 Type: Article