The Usefulness and Safety of Natural Stent in a Canine Model of Tracheal Stenosis / 결핵
Tuberculosis and Respiratory Diseases
;
: 431-438, 2002.
Article
in Korean
| WPRIM
| ID: wpr-47450
ABSTRACT
BACKGROUND:
In order to investigate the usefulness and safety of Natural Stent, we performed this study in a canine model of tracheal stenosis induced using Nd-YAG laser.METHODS:
After tracheal stenosis was induced in 12 Mongrel dogs using Nd-YAG laser, either Dumon (n=6) or Natural (n=6) stent was inserted into the trachea. To assess the degree of stent migration and mucostasis, bronchoscopy was performed every week for 4 weeks, after which all stents were removed. One week after stent removal, tracheal stenosis was evaluated by bronchoscopy.RESULTS:
The degree of stent migration was not different between the dogs with Dumon stent (3.0+/-0.8) and those with Natural (2.0+/-1.0), nor was the degree of mucostasis, at Dumon (1.7+/-0.5) and Natural Stent (1.5+/-0.6), respectively. One week after stent removal, the degree of tracheal stenosis was not different between the Dumon (1.5+/-0.5) and the Natural group (1.0+/-0.4). In addition, there was no death and the degree of tracheal stenosis remained always within the safe limit (less than 2.0) in all animals.CONCLUSION:
In a canine model of tracheal stenosis induced using Nd-YAG laser, the usefulness and safety of Natural Stent were similar to those of Dumon Stent. A clinical trial is necessary to document the usefulness and safety of Natural Stent in patients with tracheal stenosis.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Trachea
/
Tracheal Stenosis
/
Bronchoscopy
/
Stents
/
Lasers, Solid-State
Limits:
Animals
/
Humans
Language:
Korean
Journal:
Tuberculosis and Respiratory Diseases
Year:
2002
Type:
Article
Similar
MEDLINE
...
LILACS
LIS