Your browser doesn't support javascript.
loading
Development and validation of an RP-HPLC method for simultaneous determination of Ramipril and Amlodipine in tablets / 药物分析学报
Journal of Pharmaceutical Analysis ; (6): 440-446, 2013.
Article in Chinese | WPRIM | ID: wpr-475479
ABSTRACT
An RP-HPLC method for the simultaneous determination of and Amlodipine (AL)Ramipril (RP) in tablets was developed and validated by Chinese Pharmacopoeia 2010. The linearity of the proposed method was investigated in the range of 0.01-0.25 mg/mL (r2 ? 0.9998) for RP and 0.014-0.36 mg/mL (r2 ? 0.9997) for AL. The limits of detection (LOD) were 0.06μg/mL and 0.02μg/mL for RP and AL, and the limits of quantitation (LOQ) were 0.2μg/mL and 0.07μg/mL, respectively. Some major impurities and degradation products did not disturb the detection of RP and AL and the assay can thus be considered stability-indicating.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Pharmaceutical Analysis Year: 2013 Type: Article

Similar