Your browser doesn't support javascript.
loading
Evaluation of Measurement Uncertainty of 2 2 Clinical Chemistry Assays / 现代检验医学杂志
Journal of Modern Laboratory Medicine ; (4): 156-158, 2014.
Article in Chinese | WPRIM | ID: wpr-476033
ABSTRACT
Objective To calculate the measurement uncertainty of clinical chemical analytes according to the internal quality control (IQC)and external quality assessment (EQA)data in clinical laboratory.Methods Collected the IQC data from January to June 2013 and EQA data between 2012~2014 of clinical chemistry in clinical laboratory.Calculated the measure-ment uncertainty and extended uncertainty according to the Nordtest criteria.Results It was effective to evaluate the uncer-tainty using IQC and EQA data.ALP ranked the highest extended uncertainty and Na+ ranked the lowest uncertainty.The range of uncertainty varies greatly,electrolyte 4.27~18.16,enzyme 8.12~24.88,small molecular 4.88~12.44,protein and lipids 4.78~13.1.Conclusion The evaluation of clinical chemistry uncertainty by IQC and EQA data is simple and practi-cal,which is beneficial for assurancing the measurement accuracy.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Modern Laboratory Medicine Year: 2014 Type: Article