Your browser doesn't support javascript.
loading
The dosimetric characteristics of thermoluminescent dosimeter / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 696-699, 2015.
Article in Chinese | WPRIM | ID: wpr-481013
ABSTRACT
Objective To study the dosimetric characteristics of thermoluminescent dosimeter (TLD).Methods A total of 350 pieces of TLD 2000 type TLDs having the same sensitivity within ± 3.0% were selected.The TLDs were irradiated to study the repeatability and dosage effect of TLDs by using 6 MV X-rays and 125I seeds.In order to study the dose response,the TLDs were irradiated at different dosages by using 137Cs (662 keV γ-rays),125I seeds and 6 MV X-rays.In order to study the energy response,the TLDs were exposed to the same dose,from 125I seeds,137Cs and X-rays(48,65,83,118,250 keV and 6 MV).Results The maximum deviations of the repeatability were 2.7% and 4.0% for 6 MV X-rays and 125I seeds,respectively,and there was no effect of dose rate observed.The dose response of TLDs to 137Cs and 125I seeds were linear.For 6 MV X-rays,the linear response was within the range of 0.74-10 Gy and non-linear range was beyond 10 Gy.The energy response to 125I seeds,48,65,83,118,250 keV and6 MV X-rays,relative to the energy response of 137Cs,were 1.70,1.25,1.08,0.99,0.91,0.96 and 1.22,respectively.Conclusions TLD 2000 has a good repeatability and linear dose response for 137Cs,125I seeds and 6 MV X-rays with no dose rate effects,but the dose response is energy dependent.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2015 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2015 Type: Article