Determination of Selenium Species in Burdock and Panax Notoginseng Using Ultrasonic Assistant Extraction Combined with High Performance Liquid Chromatography-Inductively Coupled Plasma Mass Spectrometry / 分析化学
Chinese Journal of Analytical Chemistry
; (12): 1329-1334, 2015.
Article
in Zh
| WPRIM
| ID: wpr-481795
Responsible library:
WPRO
ABSTRACT
A high performance liquid chromatography-inductively coupled plasma mass spectrometric ( HPLC-ICP-MS) method for analysis of selenocystine (SeCys2), selenite (SeⅣ), selenomethionine (SeMet) and selenate ( SeⅥ) was described. Chromatographic separation was performed using 8 mmol/L citric acid solution (pH 5). The potentially interfering 40Ar40Ar+ at selenium masses of m/z 80 was reduced using CH4 as a reactive cell gas in the dynamic reaction cell. The selenium species was extracted using ultrasonic extraction system with a mixture of protease K and lipase. The limits of detection for SeCys2 , SeⅣ, SeMet, and SeⅥwere observed as 0. 088, 0. 033, 0. 30, and 0. 17 ng/mL, respectively. The correlation coefficients were greater than 0 . 9995 . The recoveries were observed in the range of 95 . 1%-114 . 6%. The developed method was successfully applied for the determination of selenium species in Burdock and Panax notoginseng.
Full text:
1
Index:
WPRIM
Language:
Zh
Journal:
Chinese Journal of Analytical Chemistry
Year:
2015
Type:
Article