Your browser doesn't support javascript.
loading
Clinical Observation of Automatic Ventricular Capture Management Confirmed by Paced Depolarization Integral Evoked Response Detection / 中国循环杂志
Chinese Circulation Journal ; (12): 358-361, 2016.
Article in Chinese | WPRIM | ID: wpr-486489
ABSTRACT

Objective:

To observe the automatic ventricular capture management (VCM) conifrmed by paced depolarization integral (PDI) evoked response detection via the follow-up study in patients with Zephyr5826 pacemaker implantation.

Methods:

A total of 102 relevant patients were enrolled. In order to conduct PDI calculation, pacemakers were set by bipolar sensing and bipolar pacing at immediately after implantation. VCM functions were observed at 1 day, 7 days and 1 month, 3, 6, 12 months after implantation, the ventricular threshold by VCM test and manual test were compared. The symptoms of pectoralis major stimulus, diaphragm stimulus and palpitation were observed in all follow-up patients.

Results:

There was 1 patient died by MI at 1 month after Zephyr5826 pacemaker implantation, the rest 101 patients were followed-up for 12 months. VCM function was successfully turned-on at immediately after implantation in all patients, no pectoralis major stimulus and diaphragm stimulus occurred. VCM function was turned-off in 6/101 (5.9%) patients at 7 days after implantation due to intolerable palpitation caused by daily automatic VCM, instead they received manual test at follow-up visit. The coincidence rate of ventricular thresholds between VCM test and manual test were 100%. Ventricular pacing output voltage by VCM was (0.99 ± 0.48) V,n=608. Compared with regular pacing output voltage (2.5V, 0.4ms), VCM function may save 84% of energy consumption; compared to high pacing output voltage (3.5V, 0.4ms), VCM may save 92%. Loss of ventricular capture and poor sensation were not found by ECG and 24 h dynamic monitoring.

Conclusion:

Zephyr5826 pacemaker may conduct bipolar pacing and scanning with VCM function, it can be effectively and safely operated by low energy output. A few patients may not use VCM function due to intolerable palpitation.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study / Observational study / Prognostic study Language: Chinese Journal: Chinese Circulation Journal Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study / Observational study / Prognostic study Language: Chinese Journal: Chinese Circulation Journal Year: 2016 Type: Article