Evaluation verification of Mindary BC5390 hematology analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine
;
(12): 1046-1048, 2016.
Article
in Chinese
| WPRIM
| ID: wpr-486629
ABSTRACT
Objective To evaluate the performance verification of Mindary BC5390 hematology analyzer .Methods According to requirement of CLSI documents ,the precision ,accuracy ,linearity ,carryover were evaluated .Results The background counts all met the designed requirements of manufacturer .Precision ,accuracy ,linearity of the WBC ,RBC ,PLT ,Hb ,HCT ,MCV were good and the contamination carrying rates were low .Except for basophils ,the classification of leukocyte was well correlated with the artificial microscopy classification(r2 >0 .95) .Conclusion The BC5390 is a ideal automated hematology analyzer which has high precision and accuracy ,wide linearity ,low contamination carrying rate ,and it can be used for the blood samples analysis .
Full text:
Available
Index:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
International Journal of Laboratory Medicine
Year:
2016
Type:
Article
Similar
MEDLINE
...
LILACS
LIS