Your browser doesn't support javascript.
loading
Comparison of different analyzing methods for detection of free amino acids in plasma / 中华临床营养杂志
Chinese Journal of Clinical Nutrition ; (6): 115-120, 2016.
Article in Chinese | WPRIM | ID: wpr-486904
ABSTRACT
Objective To evaluate the accuracy of different methods in detecting free amino acids in plasma.Method 40 blood samples from healthy volunteers were analyzed by an automatic amino acid analyzer ( Li+system) and the results compared with previous reports using other analyzers.Results The results ob-tained by this analyzer for major amino acids [asparagine (Asn), glutamic acid (Glu), glutamine (Gln), valine (Val), tyrosine (Tyr), phenylalanine (Phe), ornithine (Orn), arginine (Arg)] were similar with those previously reported using other amino acid analyzers ( all P>0.05 ) and liquid chromatography tandem-mass spectrometry (LC-MS/MS) (all P>0.05).Except for Glu and Tyr, the results for major amino acids showed large deviation compared to the results from high performance liquid chromatography ( HPLC) ( all P<0.05).Conclusion The amino acid analyzer (Li+) and LC-MS/MS (iTRAQ kit) could accurately detect free amino acids in plasma.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study Language: Chinese Journal: Chinese Journal of Clinical Nutrition Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study Language: Chinese Journal: Chinese Journal of Clinical Nutrition Year: 2016 Type: Article