Your browser doesn't support javascript.
loading
Discussions of Multi-Detector Computed Tomography (MDCT) dose assessment based on information in DICOM images / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 143-148, 2016.
Article in Chinese | WPRIM | ID: wpr-488585
ABSTRACT
Objective To establish an accuracy test method for MDCT dose assessment based on information in DICOM images.Methods The type of MDCT studied in this paper was widely used in clinical practice.A software package developed by java language was used to automatically read doserelated information from DICOM files of MDCT.The CTDIvol and DLP of each pectoral or abdominal scan was calculated based on these information and the basic scan parameters such as collimation,mAs and pitch.The calculated values were compared with the displayed values.Results For pectoral scans,the difference between the calculated and displayed values was between-2%-8% for CTDIvol,and-2%-5% for DLP.For abdominal scan,the difference between the calculated and displayed values was 0-2% for CTDIvol,and-2%-3% for DLP.Conclusions This method is useful for MDCT dose assessment and is worth disseminating its application for general use.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2016 Type: Article