Structure and Property of Ladder Electrode Linear Ion Trap Mass Analyzer / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 994-1000, 2016.
Article in Chinese | WPRIM | ID: wpr-494378
ABSTRACT
Theoretical study and experimental test of a new ion trap mass analyzer-ladder electrode linear ion trap (LeLIT) were reported. The LeLIT was composed of two pairs of ladder electrodes and one pair of end-electrodes. By optimizing the geometric structure of ladder electrodes, we optimized the electric field distribution for optimum mass analysis. Since the distribution of electric field can be controlled by the adjustment of LeLIT structure, LeLIT have better performance than rectilinear ion trap theoretically. Due to its simpler structure than hyperbolic electrode LIT, LeLIT can be easily built. In this work, the height, width and field ratio were adjusted for optimizing the property of LeLIT, and a mass resolution of 10150 was obtained when the mass scan speed was 225 Da / s for a LeLIT with X0 ×Y0 = 9 mm×5 mm dimension. The preliminary experimental result showed that the LeLIT has proper tandem mass spectrometric performance as any conventional ion trap mass analyzer.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Analytical Chemistry Year: 2016 Type: Article