ABSTRACT

Objective:

To evaluate the measurement uncertainty in the determination of sodium valproate tablets by GC with an in -ternal standard method , and determine the main sources of uncertainty .

Methods:

A GC internal standard method was selected to sys-tematically analyze the uncertainty in the determination of sodium valproate tablets , including the sample quantity , dilution ratio, purity and area repeatability of chromatographic peaks .

Results:

The expanded uncertainty of sodium valproate tablets was 2.7%, and the determination range of sodium valproate tablets was (96.3 ±2.7)%(k=2).

Conclusion:

The established GC internal standard meth-od for the uncertainty evaluation is reliable , which is helpful to improving the quality evaluation and control of sodium valproate tablets .