Comparison of Shikani with HC laryngoscope in nasal intubation / 实用口腔医学杂志
Journal of Practical Stomatology
;
(6): 557-560, 2016.
Article
in Chinese
| WPRIM
| ID: wpr-495310
ABSTRACT
Objective:
To analyse pros and cons of Shikani laryngoscope anatomically during nasal intubation.Methods:
60 pa-tients,with ASAⅠ -Ⅱ,aged 1 9 to 29 years old,undergoing titanium plate and titanium screw removal surgery after orthognathia, were randomly divided into Shikani group (S group,n =30)and HC video laryngoscope group (V group,n =30).Successful intuba-tion rate of the first attempt,total time cost,nasal bleeding rate and the incidence of postoperative airway complications were compared.Results:
Compared with group V,the time cost of group S was significantly lower(P 0.05).Conclusion:
Shikani la-ryngoscope is faster,less invasive with easy maintenance in nasal intubation.