Your browser doesn't support javascript.
loading
The filter processing in forced oscillation respiratory impedance measurements / 中国医学物理学杂志
Chinese Journal of Medical Physics ; (6): 49-50, 2001.
Article in Chinese | WPRIM | ID: wpr-500179
ABSTRACT
In the measurement of the forced oscillation respiratory impedance, we designed a narrow-band second-order band-pass comb filter which is centered at the forced oscillation frequencies for eliminating the interference of spontaneous breathing. The result shows that an increase of signal-to-noise ratio (SNR) reaches 6.0 dB, and keeps it almost constant over the whole frequency band when compared with a conventional high-pass filtering. So we can assess correctly the impedance of respiratory system.
Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Medical Physics Year: 2001 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Medical Physics Year: 2001 Type: Article