Study on the method to scale 192ir radioactivity / 中国医学物理学杂志
Chinese Journal of Medical Physics ; (6): 195-196, 2000.
Article in Chinese | WPRIM | ID: wpr-500259
ABSTRACT

purpose:

To calibrate the accurate value of 192Ir radioactivity again. materials and

methods:

To measure the dose rate of radioactivity in water and at air by ion chamber.

results:

To scale the activity of 192Ir radioactivity by air Kerma is agreement to the scaling method according to the JJG 773-92.

conclusion:

The investigation demonstrates that the accurate value of 192Ir radioactivity can be calibrated by the method of air Kerma.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Medical Physics Year: 2000 Type: Article