Your browser doesn't support javascript.
loading
Determination of Cu, Ni and Zn in Aluminum Food Packaging Materials by Strong Acid Cation Exchange Fiber Solid Phase Extraton-ICP-AES / 中国药师
China Pharmacist ; (12): 1965-1968, 2016.
Article in Chinese | WPRIM | ID: wpr-503255
ABSTRACT

Objective:

To establish the analysis method for the determination of Ni, Cu and Zn in aluminum food packaging mate-rials by inductively coupled plasma atomic emission spectrometry ( ICP-AES) .

Methods:

A small amount of EDTA was added to the test solution till the concentration at 0. 010 0 mol·L-1 and the pH value was adjusted to 5. 0. Strong acid cation exchange fibers were used for the assisted extraction of Al3+ with ultrasonic. Al3+ could not quickly form a complex anion with EDTA, so it was adsorbed by the strong acid cation exchange fibers. Cu2+, Ni2+ and Zn2+ could form complex anions with EDTA rapidly, so they could not be ad-sorbed by the strong acid cation exchange fibers and were left in the solution to be determined.

Results:

After the separation, the con-tents of Cu, Ni and Zn were not lost, and the content of residual aluminum was about 170 times of that of Cu, Ni and Zn, which had not interferece with the determination of Cu, Ni and Zn. The sample standard addition recovery was within the range of 98. 3%-104%with RSD of 0. 2%-2. 5%(n=5). The relative errors between the measured values and the certified values of standard substances were less than 5. 0%.

Conclusion:

The proposed method can be successfully applied in the separation and determination of Cu, Ni and Zn in aluminum cans, barbecue aluminum foil and standard substances.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Pharmacist Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Pharmacist Year: 2016 Type: Article