Your browser doesn't support javascript.
loading
A quantitative analysis of tumor site specific setup based on 1 485 daily MVCT scans from helical tomotherapy / 中国基层医药
Chinese Journal of Primary Medicine and Pharmacy ; (12): 3361-3365, 2016.
Article in Chinese | WPRIM | ID: wpr-504161
ABSTRACT
Objective To assess the variation in patient setup corrections for three different anatomic treat-ment sites using daily pretreatment megavoltage CT(MVCT)in helical tomotherapy,and to analyze alternative refer-ence margins for specific tumor site.Methods Sixty patients treated for three anatomical sites on helical tomotherapy were analyzed.Daily MVCT was carried out for registration and setup corrections before each treatment fraction.Setup errors and rotational setup corrections from 587 head and neck,500 thoracic,371 abdomen and gynecology MCVT imaging were input to statistical analysis.Incidences of three dimensional vector error lengths were calculated for each anatomic site.Overall distribution histograms of the three -dimensionally error was presented using GraphPad Prism 5.The distributions of systematic and random setup errors were quantitative analyzed and the additional margins required were also taken into account.Results 1 458 MVCT scans were implemented for sixty patients.Head and neck had lower frequencies of translational setup errors than others.Frequency of at least 3mm three -dimensional setup errors for head and neck,thoracic,abdomen and gynecology was 55.3%,70.8%,79.8%,respectively.This fre-quency decreased to 17.5%,40.6%,47.2% if 3D vector distance ≥5mm was scored.Overall systematic errors ranged from -1.4mm to 2.7mm,abdomen and gynecology had the largest setup errors in the vertical direction which was statistically significant(χ2 =19.3,P <0.05).The suggested margins should be increased by 4 -7 mm in three -dimensional direction for head and neck,9 -14mm for thoracic,10 -17mm for abdomen and gynecology.Conclusion Differences in setup corrections are perceived between head and neck,thoracic,abdomen and gynecology.The accura-cy of patient positioning can be improved if pre -treatment daily MVCT scans are put into use.Results from setup cor-rection can provide evidence for tumor treatment margin and improve the accuracy of regular radiotherapy.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Primary Medicine and Pharmacy Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Primary Medicine and Pharmacy Year: 2016 Type: Article