Comparative Stndy on Concentration Determination of Valproic Acid in Human Serum by LC-MS/MS and EMIT / 中国药房
China Pharmacy ; (12): 1049-1052, 2017.
Article in Chinese | WPRIM | ID: wpr-510099
ABSTRACT

OBJECTIVE:

To compare the difference in the concentration determination of valproie acid (VPA) in human serum by LC-MS/MS and EMIT.

METHODS:

Both LC-MS/MS and EMIT methods were applied to determine the serum concentration of VPA in 144 inpatients or outpatients.The paired t-test,Pearson correlation analysis,Bland-Altman deviation chart and other methods were used to evaluate the difference in the results of concentration determination.

RESULTS:

The results of LC-MS/MS method was pos itively correlated with that of EMIT method (r=0.924,P<0.05);the regression equation of them was cEMIT=0.920 7cLC.MS/MS-1.114 4 (r=0.924).Average serum concentrations of VPA determined by LC-MS/MS and EMIT were (49.9 ± 21.2) and (54.9 ± 21.3) μg/mL,with statistical significance (P<0.05).The serum concentration of VPA determined by EMIT was higher than that by LC-MS/MS 8.3 μg/mL,95% confidence interval was (-13.6,18.7).

CONCLUSIONS:

The serum concentration of VPA determined by LC-MS/MS and EMIT have high correlation.But the determination results still have certain difference,it is suggested to use same method for long term monitering.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Pharmacy Year: 2017 Type: Article