Content Determination of Azithromycin Suspension by HPLC / 中国药房
China Pharmacy ; (12)2007.
Article in Zh | WPRIM | ID: wpr-534376
Responsible library: WPRO
ABSTRACT
OBJECTIVE:To establish the HPLC method for content determination of Azithromycin suspension.METHODS:The determination was performed on Venusil XBP-C18(L) (150 mm?4.6 mm,5?m) column with mobile phase consisted of 0.05 mol?L-1 dipotassium phosphate (pH value adjusted to 8.2 by 20% phosphoric acid)-acetonitrile (45:55) at flow rate of 1.2 mL?min-1.Injection volume was 20 ?L and UV detection wavelength was set at 210 nm.RESULTS:The good resolution between peaks of azithromycin and that of adjacent purity was more than 1.5.The linear range of azithromycin was 0.01~5 mg?mL-1 (r= 0.999 9) with an average recovery of 98.83%(RSD=0.54%,n=6).The precision of intra-day,inter-day and with-day were all lower than 2.5%.CONCLUSION:The method is specific and accurate for the content determination of Azithromycin suspension.
Key words
Full text: 1 Index: WPRIM Language: Zh Journal: China Pharmacy Year: 2007 Type: Article