An assessment of hydrodynamics of the C-L pugestrut valve prosthesis in vitro / 第二军医大学学报
Academic Journal of Second Military Medical University
;
(12)1999.
Article
in Chinese
| WPRIM
| ID: wpr-551641
ABSTRACT
Objective:
To assess the hydrodynamic performance of C L pugestrut valve prosthesis in vitro .Methods:
The leakage, steady flow and pulsatile flow of the valve prosthesis was tested in vitro with the C L standard valve and Medtronic Hall valve prosthesis as control.Results:
The rate of leakage of the C L pugestrut valve was (298?66) ml/min, accorded with the demand of clinical use. The steady flow pressure drop across the C L pugestrut valve was significantly lower than the C L standard valve( P
Full text:
Available
Index:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Academic Journal of Second Military Medical University
Year:
1999
Type:
Article
Similar
MEDLINE
...
LILACS
LIS