Biomechanical analysis of the failure of the interlocking nail of femal / 临床外科杂志
Journal of Clinical Surgery
;
(12)2001.
Article
in Chinese
| WPRIM
| ID: wpr-553675
ABSTRACT
Objective To evaluate the reason of failure of the interlocking nail.Methods From1998-2001, 72 patients with femur fractures were treated with the interlocking nail.Results 61 cases were followed-up,there were 5 breakages in the locking nail. The failure occurred after operation from 4~9 months.And 3 cases were at the middle of the nail.2 cases were at the first screw of the distal of the fracture.Conclusion The reason of breakage concluded open operation,too much separation of the periosteum,too early full weight bearing and unfit function exercise.