Your browser doesn't support javascript.
loading
The feasibility of direct applying MRI to treatment planning system of radiotherapy for intra-cranial tumor / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6)1992.
Article in Chinese | WPRIM | ID: wpr-556904
ABSTRACT
Objective Objective To study the feasibility of direct applying MRI to TPS by measuring the geometric distortion of MRI. Methods Phantom 1 was made by marking radical arrangement of fat pipelines every 15? in foam phantom and Phantom 2 by net arrangement at every 2.7?cm. Coronal and transverse scan parallel with the radial and net fat pipeline in the phantom was performed at sequence of cranial FSE T2WI by Marconi 1.5T superconduct MRI machine. Geometric distortion in 5 different fields of view (FOV) was calculated by measuring the absolute borderline error of direct scan and coronal digital reconstructed radiograph (DRR). Results When FOV was equal or smaller than 11.0?cm?11.0?cm, 22.0?cm?22.0?cm and 27.5?cm?27.5?cm, the geometric distortion was 0.0, 1.1 and 1.4?mm of direct scanning MRI and 1.0,1.4 and 2.0?mm of coronal DRR. Conclusion When the magnetic field of MR is high and the tissue shows uniform density, MRI can be applied on TPS directly.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 1992 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiation Oncology Year: 1992 Type: Article