Study on piezoelectric quartz crystal microarray immunosensor for the detection of Agkistrodon acutus venom / 第三军医大学学报
Journal of Third Military Medical University
;
(24)2002.
Article
in Chinese
| WPRIM
| ID: wpr-565862
ABSTRACT
Objective To explore the possibility of piezoelectric microarray immunosensor for the detection of Agkistrodon acutus venom. Methods Microarray immunosensor with quartz crystal of 10 MHz AT-cut and 2?5 gold-coated electrodes was prepared. The thiol-treated venom antibody was immobilized by a self assembling device for the detection of the standard fluid for different concentrations of the venom. Results Experimental results showed that the optimal concentration of the antibody was 3.0 g/ml with the response time of 40 minutes. The piezoelectric immunosensor could well respond to homologous venoms. Within the range of 0.1~4.0 g/ml, the frequency shifts were linearly dependent on the venom concentration. Conclusion Piezoelectric microarray immunosensor for the detection of Agkistrodon acutus venom is of high specificity of response, high sensitivity, and simple operation without marking. The technique of piezoelectric microarray immunosensor is possible to test snakebite quickly, quantitatively, and instrumentally.
Full text:
Available
Index:
WPRIM (Western Pacific)
Type of study:
Diagnostic study
Language:
Chinese
Journal:
Journal of Third Military Medical University
Year:
2002
Type:
Article
Similar
MEDLINE
...
LILACS
LIS